Inter, intra-examiner reliability and validity of inertial sensors to measure the active cervical range of motion in patients with primary headache
DOI:
https://doi.org/10.17179/excli2021-3799Keywords:
primary headache, reliability, validity, physical examination, range of motionAbstract
We analyzed the inter- and intra-examiner reliability of Werium inertial sensors and the cervical range of motion (CROM) instrument for the measurement of active CROM (AcROM) in patients with primary headache. Another objective is to analyze the validity of the inertial sensors (Werium). The literature has reported symptomatology features in patients diagnosed with primary headache similar to that of patients with cervicogenic headache. The International Classification of Headache (ICHD-III) established the presence of reduced AcROM as a diagnostic criterion for cervicogenic headache. Several instruments are used for this measurement, with limitations in their applicability in daily clinical practice. A prospective longitudinal repeated measures study was conducted to assess the intra- and inter-rater reliability and validity of Werium inertial sensors in 20 adults with chronic primary headache. For the inter-rater analysis, the intraclass correlation coefficient (ICC) values were above 0.75 for all movements, indicating a good level of reliability. For the intra-rater results, the ICC values obtained by the Werium inertial sensors for all cervical movements were good for rater A (ICC >0.80) and rater B (ICC >0.84). For the validity, the ICCs obtained by the Werium inertial sensors compared with the CROM instrument for all cervical movements were moderate for both raters (ICC > 0.70, respectively). Values obtained in the standard error of measurement, minimum detectable change at 90% and limits of agreement also indicated good agreement. Werium inertial sensors have shown good to excellent reliability results, both intra- and inter-examiner (ICC > 0.75). Likewise, when the sensors were compared with another validated instrument (CROM device) they obtained high reliability results (ICC > 0.70). These results plus its relatively low price and ease of use allow us to recommend it in daily clinical practice to measure AcROM in patients with chronic primary headache.
Downloads
Additional Files
Published
How to Cite
Issue
Section
License
Copyright (c) 2021 Ignacio Elizagaray-García, Alfonso Gil-Martínez, Gonzalo Navarro-Fernández, Ana Rosa Navarro-Moreno, Jacqueline Sánchez-de-Toro-Hernández, Javier Díaz-de-Terán, Sergio Lerma-Lara
This work is licensed under a Creative Commons Attribution 4.0 International License.
Authors who publish in this journal agree to the following terms:
- The authors keep the copyright and grant the journal the right of first publication under the terms of the Creative Commons Attribution license, CC BY 4.0. This licencse permits unrestricted use, distribution and reproduction in any medium, provided that the original work is properly cited.
- The use of general descriptive names, trade names, trademarks, and so forth in this publication, even if not specifically identified, does not imply that these names are not protected by the relevant laws and regulations.
- Because the advice and information in this journal are believed to be true and accurate at the time of publication, neither the authors, the editors, nor the publisher accept any legal responsibility for any errors or omissions presented in the publication. The publisher makes no guarantee, express or implied, with respect to the material contained herein.
- The authors can enter into additional contracts for the non-exclusive distribution of the journal's published version by citing the initial publication in this journal (e.g. publishing in an institutional repository or in a book).